Tools for Functional Imaging > MRI-Compatible Displays > BOLDscreen for fMRI
Designed specifically for fMRI, BOLDscreen offers high brightness, high contrast and unparallelled stimulus control.
It is ideal for visual stimulation and patient entertainment.
System Environmental Conditions
Indoor use only.
Operating Temperature: 0°C to +30°C
Cooling : Convection-cooled
Operating (or Storage Humidity up to 40°C) : 10-90% RH, non-condensing
Storage Temperature: -20 °C to +60 °C, 40% RH.
Monitor Characteristics
Brightness: maximum 1000cd m-2
The normal switching frequency of the backlight is 300Hz.
Typical uniformity 2% over central 75% of display area.
Automatic compensation of brightness for temperature and ageing (factory set to maintain 800cd m-2).
Grey-to-Grey Response time: 5ms typical
Video Input: 1920x1200 pixels, DVI, 60Hz, 24 bit.
Backlight trigger: Opto isolated TTL input 0 to 5Vdc max.
RF emissions: No effect observed on functional and structural MRI scans at 1.5 and 3T with screen located at exit of scanner bore. No effect on displayed image whilst scanning.
Type B equipment in accordance with EN60601 for devices in the patient environment.
Monitor Dimensions
Dimensions: 620 x 445 x 70mm
Weight: 16.5kg
Stand Dimensions
Height adjustable: 1000 – 1800mm
Weight: 2.5kg
Power Supply
Mains Input: 100 to 240Vac nominal, 50 or 60Hz, 150w.
Output: 5, 15 and 35 to 45Vdc.
Fuses: HBC T3.15A / 250VAC
Dimensions: 282 x 187 x 75mm
Weight: 3.75 kg
EMC
Emissions: EN55011 class B conducted & radiated
Harmonic Currents : EN61000-3-2 class A
Voltage Flicker: EN61000-3-3
ESD Immunity: EN61000-4-2, level 3 contact, Perf Criteria A
Radiated Immunity: EN61000-4-3, level 3, Perf Criteria A
EFT/Burst: EN61000-4-4, level 2, Perf Criteria A
Surge: EN61000-4-5, installation class 3,Perf Criteria A
Conducted Immunity: EN61000-4-6, level 3, Perf Criteria A
Magnetic Field
(limit applies to PSU only): EN61000-4-8, 3 A/m, Perf Criteria A
Dips & Interruptions : EN61000-4-11, 30% 10 ms, 60% 100 ms, >95% 5000 ms
Perf Criteria A, B, B
Safety Conformance
EN60601-1, Type B Equipment, Class 1
Cables
Optical Fibre Length: 20m
DC Power Cable Length: 10m
Backlight trigger cable: BNC to SMA, 10m.
Mains power lead: IEC
USB cable: Type A-B, length: 1.5m