BOLDscreen for fMRI

boldscreen-with-fish.png

Designed specifically for fMRI, BOLDscreen offers high brightness, high contrast and unparallelled stimulus control.

It is ideal for visual stimulation and patient entertainment.

System Environmental Conditions

Indoor use only.

Operating Temperature: 0°C to +30°C

Cooling : Convection-cooled

Operating (or Storage Humidity up to 40°C) : 10-90% RH, non-condensing

Storage Temperature: -20 °C to +60 °C, 40% RH.

 

Monitor Characteristics

Brightness: maximum 1000cd m-2

The normal switching frequency of the backlight is 300Hz.

Typical uniformity 2% over central 75% of display area.

Automatic compensation of brightness for temperature and ageing (factory set to maintain 800cd m-2).

Grey-to-Grey Response time: 5ms typical

Video Input: 1920x1200 pixels, DVI, 60Hz, 24 bit.

Backlight trigger: Opto isolated TTL input 0 to 5Vdc max.

RF emissions: No effect observed on functional and structural MRI scans at 1.5 and 3T with screen located at exit of scanner bore. No effect on displayed image whilst scanning.

Type B equipment in accordance with EN60601 for devices in the patient environment.

 

Monitor Dimensions

Dimensions: 620 x 445 x 70mm

Weight: 16.5kg

 

Stand Dimensions

Height adjustable: 1000 – 1800mm

Weight: 2.5kg

 

Power Supply

Mains Input: 100 to 240Vac nominal, 50 or 60Hz, 150w.

Output: 5, 15 and 35 to 45Vdc.

Fuses: HBC T3.15A / 250VAC

Dimensions: 282 x 187 x 75mm

Weight: 3.75 kg

 

EMC

Emissions: EN55011 class B conducted & radiated

Harmonic Currents : EN61000-3-2 class A

Voltage Flicker: EN61000-3-3

ESD Immunity: EN61000-4-2, level 3 contact, Perf Criteria A

Radiated Immunity: EN61000-4-3, level 3, Perf Criteria A

EFT/Burst: EN61000-4-4, level 2, Perf Criteria A

Surge: EN61000-4-5, installation class 3,Perf Criteria A

Conducted Immunity: EN61000-4-6, level 3, Perf Criteria A

Magnetic Field

(limit applies to PSU only): EN61000-4-8, 3 A/m, Perf Criteria A

Dips & Interruptions : EN61000-4-11, 30% 10 ms, 60% 100 ms, >95% 5000 ms

Perf Criteria A, B, B

 

Safety Conformance

EN60601-1, Type B Equipment, Class 1

 

Cables

Optical Fibre Length: 20m

DC Power Cable Length: 10m

Backlight trigger cable: BNC to SMA, 10m.

Mains power lead: IEC

USB cable: Type A-B, length: 1.5m

 

Other products you might be interested in

LiveTrack AV for fMRI

Affordable MRI-compatible video eye tracking with real-time estimates of eye...

Click to read more »

BrainVoyager QX

Analysis and visualization of functional and anatomical magnetic resonance imaging...

Click to read more »

MediGoggle Adult Research Set

Interchangeable prescriptive goggles suitable for use in MRI, fMRI, MEG,...

Click to read more »